Using The Manual Charge Feature (M3860A Only) - Philips M3860A Instructions For Use Manual

Defibrilator
Table of Contents

Advertisement

6-11
In the M3861A HeartStart FR2+, the patient's ECG is not displayed; in the
M3860A, the display includes the patient's ECG and heart rate.
Press the lower Option button (
) to initiate rhythm analysis by the
ANALYZE
FR2+. If a shock is advised, the FR2+ automatically charges, and prompts
you to press the Shock button.
After shock delivery, the HeartStart FR2+ returns to the advanced mode
display and monitors the patient's heart rhythm. If a potentially shockable
rhythm is detected, the text and voice prompts advise you to
PRESS
.
ANALYZE
:
If you do not press the lower Option button (labeled
) to
N O T E
ANALYZE
initiate rhythm analysis when prompted, the HeartStart FR2+ does not
analyze and prompt if a shock is advised. It is important that you understand
that entering the advanced mode entails taking responsibility for these
functions.
If the rhythm analysis results in a Shock Advised decision, the FR2+ begins
charging, prompts you to press the Shock button, and displays a
MANUAL
option at the top of the screen. If for any reason you want to cancel
DISARM
the shock, press the upper Option button to disarm the FR2+.
To return to non-manual, AED mode operation, turn the FR2+ off by pressing
the On/Off button. Then turn the FR2+ on by pressing the On/Off button
again.

Using the manual charge feature (M3860A only)

6
The manual charge feature is available only in the M3860A, when enabled in
setup.
To enter the advanced mode during use of an FR2+ that has this feature
enabled, press both Option buttons simultaneously. This brings up a screen
that includes a highlighted line at the top, labeled
, with an
MANUAL
arrowhead pointing to the upper Option button, and another at the bottom,
labeled
, with an arrowhead pointing to the lower Option button.
ANALYZE
When the advanced mode is entered, display of the patient's ECG and heart
rate is automatically initiated.
Pressing the lower Option button (
) provides user-initiated rhythm
ANALYZE
analysis as described above. Pressing the upper Option button (
)
MANUAL
brings up a new screen.
S et u p a n d A d v an c e d M o d e F e a tu r e s

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

M3861aHeartstart fr2+

Table of Contents