Philips M3860A Instructions For Use Manual page 77

Defibrilator
Table of Contents

Advertisement

category
Application
Length and Weight
Operating Temperature and
Humidity
Storage Temperature and
Humidity
Patient Lead Wire
Designation
Typical (Lead II) Connection
Battery Type
Service Life
Performance with FR2+
Defibrillator
(Optional) M3873A/M3874A FR2+ ECG assessment
module
nominal specifications
For use with the FR2+ M3860A with ECG display enabled and running version
1.5 software or higher (denoted by FR2+ on the front panel or rear label).
100 inches (182 cm); ≤ 1 lb. (2.2 kg).
32° to 122° F (0° to 50° C);
0% to 95% relative humidity (non-condensing).
32° to 109° F (0° to 43° C).
0% to 75% relative humidity (non-condensing).
M3873A (AAMI):
positive lead − red
negative lead − white
reference lead − black.
Lead II vectors:
positive − left leg
negative − right arm
reference − left arm.
Other limb vectors can be obtained by different electrode positions.
3 V, 1 Ah, poly-carbonmonofluoride lithium (LiCFx). Non-replaceable disposable
primary cell.
Typically, 5 years.
Meets environmental specifications cited for FR2+ Defibrillator on page B1
through B2.
M3874A (IEC):
positive lead − green
negative lead − red
reference lead − yellow
Te ch n ic al S p ec if i ca ti o n s
B-9
B