After Using the HeartStart FR2+

After each use of the FR2+, perform the maintenance tasks described in the
table above, as well as the following post-use checks before returning the
FR2+ to service:
Check the operation of the FR2+ by removing and reinstalling the
battery and running the battery insertion selftest.
when replacing expired defibrillator pads.
Check the outside of the FR2+ and the connector socket for signs of
dirt or contamination. If the FR2+ is dirty or contaminated, clean it
according to the guidelines provided in this manual.
Check the data card if one has been used. If the data card has been
used to record incident data, remove and replace it with a blank data
card. Deliver the recorded data card to appropriate personnel according
to your local guidelines and medical protocol.
maintenance task/response
Check the Status Indicator.
If you see the flashing black hourglass: The FR2+ is
ready to use. No action required.
If you see anything other than a flashing black
hourglass, remove and reinstall the battery to run the
selftest.
• If the selftest passes and the Status Indicator
shows the flashing black hourglass, the FR2+ is
ready to use.
• If the selftest fails, install a new battery and run the
selftest. If the selftest passes, the FR2+ is ready to
use. If the selftest fails, contact Philips Medical
Systems.
Check supplies, accessories, and spares for
damage and expiration dating.
Do not use damaged or expired accessories. Replace
them immediately.
If a
or
LOW BATTERY
REPLACE BATTERY
displayed: Replace the battery and run the selftests.
DO NOT ATTEMPT TO CHARGE THE
. It is not rechargeable. The
STANDARD BATTERY
M3848A FR2+ battery is rechargeable. Recharge it,
using the M3849A Charger, for the FR2+
rechargeable battery only.
Check the outside of the FR2+ and the connector
socket for cracks or other signs of damage.
If you see signs of damage: Contact Philips Medical
Systems for technical support.
message is
M3863A FR2
:
Perform also
N O T E