uninterrupted view of the "eye" in the other. (See the diagram on the
back of the first page of this manual.) The two devices should be no
more than 1 meter apart.
5. Select
from the
SEND TIME
FR2+.
6. Select
RECEIVE TIME
7.
A new screen comes up. Until the two FR2+ devices are properly
positioned, the screen on the receiving FR2+ displays
and prompts you to check the sending FR2+. The screen on
RECEIVE
the sending FR2+ displays
the receiving FR2+.
8. Clock settings are automatically transferred as soon as the infrared ports
are correctly aligned.
Using Advanced Mode Features
The HeartStart FR2+ provides an advanced mode that allows responders
who are appropriately trained to override the programmed FR2+ protocol
and take responsibility for certain aspects of the operating sequence used by
the FR2+ to treat the patient.
As described earlier in this chapter, the factory default setup of the FR2+
must be modified to provide access to advanced mode features. This
requires use of the administration function of the M3864A training &
administration pack.
If you are an expert user authorized by your Medical Director to modify setup,
hold down both the Option buttons while installing the training &
administration pack in the FR2+, then select
from the
menu. Select
SETUP
SETUP
menu.
MODIFY SETUP
Using the upper Option button, scroll through the available settings for
. The advanced mode options available are based on the FR2+
ADVANCED
screen on the "sending" HeartStart
CLOCK
from the
screen of the "receiving" FR2+.
CLOCK
and prompts you to check
READY TO SEND
SETUP
from the third menu of the
ADVANCED
S et u p a n d A d v an c e d M o d e F e a tu r e s
6-9
READY TO
. Then select
MODIFY
6