Using Advanced Mode Features - Philips M3860A Instructions For Use Manual

Defibrilator
Table of Contents

Advertisement

uninterrupted view of the "eye" in the other. (See the diagram on the
back of the first page of this manual.) The two devices should be no
more than 1 meter apart.
5. Select
from the
SEND TIME
FR2+.
6. Select
RECEIVE TIME
7.
A new screen comes up. Until the two FR2+ devices are properly
positioned, the screen on the receiving FR2+ displays
and prompts you to check the sending FR2+. The screen on
RECEIVE
the sending FR2+ displays
the receiving FR2+.
8. Clock settings are automatically transferred as soon as the infrared ports
are correctly aligned.

Using Advanced Mode Features

The HeartStart FR2+ provides an advanced mode that allows responders
who are appropriately trained to override the programmed FR2+ protocol
and take responsibility for certain aspects of the operating sequence used by
the FR2+ to treat the patient.
As described earlier in this chapter, the factory default setup of the FR2+
must be modified to provide access to advanced mode features. This
requires use of the administration function of the M3864A training &
administration pack.
If you are an expert user authorized by your Medical Director to modify setup,
hold down both the Option buttons while installing the training &
administration pack in the FR2+, then select
from the
menu. Select
SETUP
SETUP
menu.
MODIFY SETUP
Using the upper Option button, scroll through the available settings for
. The advanced mode options available are based on the FR2+
ADVANCED
screen on the "sending" HeartStart
CLOCK
from the
screen of the "receiving" FR2+.
CLOCK
and prompts you to check
READY TO SEND
SETUP
from the third menu of the
ADVANCED
S et u p a n d A d v an c e d M o d e F e a tu r e s
6-9
READY TO
. Then select
MODIFY
6