3. Press the upper Option button to activate the test.
4.
N O T E
the FR2+ during a battery insertion selftest, the selftest will stop and the
FR2+ will go to its standby mode to be ready for use.
When the automatic part of the selftest is successfully completed, the screen
displays a message that the test has passed, and then automatically starts
the interactive part of the selftest. It is important to press the buttons and
verify the indicators to ensure that the FR2+ will be ready for use.
GOOD BATTERY
When the entire selftest is complete, the FR2+ automatically turns off and
returns to standby mode. In the standby mode, the Status Indicator displays
a flashing black hourglass. This means that the FR2+ has passed its most
recent self-test and is therefore ready for use, simply by pressing the On/Off
button to turn it on.

Placing and Securing the HeartStart FR2+

Place the HeartStart FR2+ Defibrillator in an accessible area with the Status
Indicator easily visible. Useful accessories for placing and securing the FR2+
include a carrying case, which is suitable for use with a wall mount bracket or
defibrillator cabinet. (See Appendix A for a list of accessories.)
N O T E
open the pads package until ready for use.
With the battery installed and the FR2+ stored in appropriate environmental
conditions, the FR2+ performs detailed periodic selftests to make sure that it
remains ready for use. (See Appendix B for the environmental storage
specifications.)
While the FR2+ is in the standby mode, the Status Indicator shows the
flashing black hourglass unless the periodic selftests detect a problem. If a
problem is detected, the Status Indicator will show a flashing red X or a solid
red X and the FR2+ will beep ("chirp") to alert you to the need for
troubleshooting. (See Chapter 4, "Maintaining, Testing, and Troubleshooting
Your HeartStart FR2+, " for instructions.)
make no selection for 10 seconds, and the selftest will start
OR
automatically if the FR2+ has been turned off for at least 5 minutes.
:
If you connect defibrillator pads (that are applied to the patient) to
:
Do not store the FR2+ with the defibrillator pads attached. Do not
P r ep a r in g Yo u r H ea r tS t a r t F R 2+ f o r Us e
2-5