category
Battery Type
Capacity
Status Indicators
Storage Temperature
Standby Life
(after installation)
category
Application
Power Requirements
Environmental
Requirements
Conformance Testing
category
Pads, Cable, and Connector
(Optional) M3848A FR2+ rechargeable battery
11.3 VDC, 6.5 Ah, lithium ion. Rechargeable cell using the M3849A charger.
When freshly charged and used at 77° F (25° C) , provides a minimum of 80
(typically 100) shocks, or 3.5 hours (typically 5 hours) of ECG display time only,
before recharging is indicated.
Good battery: bar graph on display screen indicating remaining power level.
Low battery: flashing red X on the front panel of the FR2+ (When low battery
indicator appears, there is still enough energy to deliver 9 shocks plus 15 minutes
of ECG display time).
Dead battery: solid red X on the front panel of the FR2+.
32° to 109° F (0° to 43° C).
6 months when installed fully charged in a defibrillator labeled FR2+.
(Optional) M3849A charger
nominal specifications
For use with M3848A FR2+ rechargeable battery only.
100 to 240 VAC, 47 to 63 Hz, 30 Watts
32° to 122° F (0° to 50° C).
0% to 95% relative humidity (non-condensing).
International: EN60335-1:1994 Class 1
North America: UL 1310 Class 2
M3870A and DP2/DP6 defibrillator pads
nominal specifications
Disposable and self-adhesive. DP2/DP6 adult defibrillator pads have a nominal
active surface area of 100 cm
integrated 122 cm (48 inch), typical, cable and connector. M3870A FR2
infant/child reduced-energy defibrillator pads have a nominal active surface area of
2
44 cm
each and are provided in a sealed package with an integrated 122 cm (48
inch), typical, cable and connector incorporating attenuating electronics.
nominal specifications
2
each and are provided in a sealed package with an
Te ch n ic al S p ec if i ca ti o n s