Philips M3860A Instructions For Use Manual page 76

Defibrilator
Table of Contents

Advertisement

B-8
category
Defibrillator Pad
Requirements
category
Capacity
category
Battery Type
Capacity
Status Indicators
Storage Temperature
category
Application
Power Requirements
Environmental
Requirements
Conformance Testing
M3860A/M3861A HEARTSTART FR2+
Use only DP2/DP6, M3870A, M3713A, and M3716A defibrillator pads with the
HeartStart FR2+. Place the pads on the patient as illustrated on each pad.
(Optional) M3854A data card
8 hours of event and ECG data, or 60 minutes with voice recording.
(Optional) M3864A training & administration pack
12 V, 1.1 Ah, nickel metal hydride. Disposable, rechargeable cell using the
M3855A charger.
Provides 4 hours of operating time at 77 °F (25 °C).
Low battery: flashing red X on the front panel of the FR2+.
Dead battery: solid red X on the front panel of the FR2+.
50° to 104° F (10° to 40° C).
(Optional) M3855A charger
For use with M3864A training & administration pack only.
With appropriate power cord, any AC mains power input or inverter-type power
sources.
32° to 113° F (0° to 45° C).
35% to 85% relative humidity (non-condensing).
International: EN60335-1:1994 Class I
North America: UL 1310 Class 2
nominal specifications
nominal specifications
nominal specifications
nominal specifications