Application Programs; Counting - Sartorius Entris Operating Instructions Manual

Laboratory balances
Table of Contents

Advertisement

Application Programs

Counting

Z
Display symbol:
Purpose
With the Counting application, you can
determine the number of parts that each
have approximately equal weight. To
do this, a known number of parts (the
reference sample quantity) is weighed first,
and the individual piece weight (reference
weight) is calculated from this result. Thus
the number of parts subsequently placed
on the balance can be determined from
their weight.
Changing the Reference Sample Quantity
Activate function:
Press the b key
Select the desired reference sample
quantity (1 to 100):
In increments of 1: Press the b key briefly
In increments of 10:
Press and hold the b key.
The quantity is stored in battery-backed
memory.
Reference Sample Updating
Automatic reference sample updating
optimizes the counting accuracy. You can
activate or deactivate this function in the
menu.
Automatic reference sample updating
is performed when the requirements,
including the specified stability criterion,
have been met.
The abbreviation opt for "optimizing", is
displayed briefly with the new reference
sample quantity.
Preparation
Select the Counting application in the
t
menu:
see "Configuration"
Set the following parameters:
t
Applic.ation programs
Count.
Resolut.ion
o Disp.Acc. Display accuracy
10-fold 10-fold higher
Ref.Updt. A utom. ref. sample
updating
o Off
autom.
o = Factory setting
Printout: Counting
nRef
10
wRef
21.14 g
Qnt
+
500 pcs
Off
Automatic
: Ref. sample quantity
: Reference weight
for 1 unit
: Calculated quantity
23

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents