Resolution; Data Processing; Fig. 141 Calculated Maximum Relative Measuring Error Relative To The Current Measured Value In The Re- Spective Measuring Range; Fig. 142 Measurement Sequence - Beckhoff EL3692 Documentation

2 channel resistance measurement terminal
Hide thumbs Also See for EL3692:
Table of Contents

Advertisement

Fig. 141: Calculated maximum relative measuring error relative to the current measured value in the
respective measuring range
Example from above: 50 Ω -> max. error 0.5 Ω, corresponding to 1% relative measuring error. The grey area
is only shown for illustration and is not a valid measuring range.
5.7.6

Resolution

The following table shows the resolution for each measuring range.
Measuring range
-1 (0xFF/255)
0
1
2
3
4
5
6
7
5.8

Data processing

Table of contents
Measurement sequence [} 120]
Processing sequence [} 121]
EL3692
Measured variable
10 m Ω - 0.1 Ω
100 m Ω - 1 Ω
1 Ω - 10 Ω
10 Ω - 100 Ω
100 Ω - 1 kΩ
1 kΩ - 10 kΩ
10 kΩ - 100 kΩ
100 kΩ - 1 MΩ
1 MΩ - 10 MΩ
Version: 2.1
Resolution
11.9 nΩ/Digit
119 nΩ/Digit
1.19 µΩ/Digit
11.9 µΩ/Digit
119 µΩ/Digit
1.19 mΩ/Digit
11.9 mΩ/Digit
119 mΩ/Digit
1.19 Ω/Digit
Commissioning
119