Interacoustics AT235 Instructions For Use Manual page 59

Impedance audiometer
Hide thumbs Also See for AT235:
Table of Contents

Advertisement

Available languages

Available languages

AT235 Instructions for Use - EN
Coupler Types used for Calibration
IMP:
TDH39 and DD45 is calibrated using a 6cc acoustic coupler made in accordance to IEC 60318-3, Ipsilateral
earphone and probe tone are calibrated using a 2cc acoustic coupler made in accordance to IEC 60318-5.
General Information about Specifications
Interacoustics continuously strives to improve its products and their performance. Therefore the
specifications can be subject to change without notice.
The performance and specifications of the instrument can only be guaranteed if it is subject to technical
maintenance at least once per year. This should be carried out by a workshop authorized by Interacoustics.
Interacoustics puts diagrams and service manuals at the disposal of authorized service companies.
Enquiries about representatives and products may be sent to:
Interacoustics A/S
Audiometer Allé 1
5500 Middelfart
Denmark
Phone:
+45 63713555
Fax:
+45 63713522
E-mail:
info@interacoustics.com
http:
www.interacoustics.com