1. Never insert or in any way use the insert headset without a new clean non defect test tip. Always make
sure that foam or ear-tip is mounted correctly. Ear tips and foam are single use.
2. The instrument is not intended to be used in environments exposed to fluid spills.
3. The instrument is not intended to be used in oxygen rich environments or use in conjunction with
flammable agents.
4. Check calibration if any parts of the equipment is exposed to shock or rough handling.
NOTICE
1. To prevent system faults take appropriate precautions to avoid PC viruses and similar.
2. Use only transducers calibrated with actual instrument. To identify a valid calibration, the serial number
for the instrument will be marked on the transducer.
3. Although the instrument fulfils the relevant EMC requirements precautions should be taken to avoid
unnecessary exposure to electromagnetic fields, e.g. from mobile phones etc. If the device is used
adjacent to other equipment it must be observed that no mutual disturbance appears. Please also refer
to EMC considerations in section 5.2
4. Use of accessories, transducers and cables other than specified, with the exception of transducers and
cables sold by Interacoustics or representatives may result in increased emission or decreased immunity
of the equipment. For list of accessories, transducers and cables that fulfils the requirements please
refer to section 5.2
5. Within the European Union it is illegal to dispose electric and electronic waste as unsorted municipal
waste. Electric and electronic waste may contain hazardous substances and therefore has to
be collected separately. Such products will be marked with the crossed-out wheeled bin shown
below. The cooperation of the user is important in order to ensure a high level of reuse and
recycling of electric and electronic waste. Failing to recycle such waste products in an
appropriate way may endanger the environment and consequently the health of human
beings.
6. Outside the European Union, local regulations should be followed when disposing of the product after its
useful life.