Interacoustics AT235 Instructions For Use Manual page 58

Impedance audiometer
Hide thumbs Also See for AT235:
Table of Contents

Advertisement

Available languages

Available languages

AT235 Instructions for Use - EN
Table 3: Reference Values for Stimulus Calibration
Freq.
Reference Equivalent Threshold Sound Level
(RETSPL)
[dB re. 20 µPa]
[Hz]
TDH39
EARtone
3A / IP30
125
45
26
250
25.5
14
500
11.5
5.5
1000
7
0
1500
6.5
2
2000
9
3
3000
10
3.5
4000
9.5
5.5
6000
15.5
2
8000
13
0
WB
-8
-5
LP
-6
-7
HP
-10
-8
*All figures in bold are Interacoustics Standard values.
CIR55
DD45 Probe NB
Stimulus
Correction
Values
26
47.5 41
4
14
27
24.5
4
5.5
13
9.5
4
0
6
6.5
6
2
8
5
6
3
8
12
6
3.5
8
11
6
5.5
9
3.5
5
2
20.5 3
5
0
12
-5
5
-5
-8
-5
-7
-6
-7
-8
-10
-8
Variation of Ipsi
Sound
stimulus levels
attenuation
for different
values for
volumes of the
TDH39/DD45
ear canal
earphones
using MX41/AR
Relative to the
or PN51
calibration
cushion
performed on
an IEC 126
[dB]
coupler
[dB]
0.5 ml
1 ml
3
5
9.7
5.3
7
9.7
5.3
15
21 (1600 Hz)
11.7
3.9
26
-0.8
-0.5
31 (3150 Hz)
-1.6
-0.8
32
26 (6300 Hz)
24
7.5
3.2
8.0
3.6
3.9
1.4