ERROR CHECKING

During weighing the balance is constantly checking to see if it is operating within the
limited parameters. The errors likely to occur are:
A/D counts below lowest allowed value
A/D counts above highest allowed value
A/D not operating
Maximum capacity exceeded
Other errors may be detected during special functions or operations. These will be
described in the section that applies.
Error messages and the reasons are:
Concerning A/D counts
ERR UL
ERR UL
ERR UL
ERR UL
ERR OL
ERR OL
ERR OL
ERR OL
Concerning calibration
ERR STB
ERR STB
ERR STB
ERR STB
ERR LO or ERR HI
ERR LO
ERR HI
ERR LO
ERR LO
ERR HI
ERR HI
Concerning weighing
ERR LO
ERR LO
ERR LO
ERR LO
ERR HI
ERR HI
ERR HI
ERR HI
© Adam Equipment Company 2015
A/D counts below a limit
A/D counts above a pre-set limit
Calibration could not be completed
because the results were not stable
Calibration constant not within 20% of old
calibration constant
Weight display is below zero by >4%max
Weight is above maximum plus 90d
37
3016612482 Rev1-Aug15