Calibration; Manual Calibration; Calibration Using Internal Calibration Mass (If Fitted); Calibration Using External Calibration Mass - Adam Equipment Eclipse EBL Series Operating Manual

Table of Contents

Advertisement

Available languages

Available languages

EN
6.2

CALIBRATION

Units with an 'i' suffix after the model name can be calibrated using either internal
calibration mechanism or by using an external mass. Units with an 'e' suffix can only be
calibrated with an external mass. Internal calibration option must be enabled in the setup
menu options or else external calibration mode will be used when the [Cal] key is
pressed.

6.2.1 Manual Calibration

Pressing the [Cal] key will start calibration. Calibration can also be initiated by a change
in internal temperature or a set time period as determined by the user.
Pressing [
0/T
 ] will abort the calibration at any time.
 
Calibration should be performed carefully and in conditions of no vibration, air movement
or other disturbance. Make sure the pan is empty, clean, and correctly fitted.

6.2.2 Calibration using Internal Calibration mass (if fitted)

Note: Internal calibration (if fitted) will only initiate if it is enabled as the default calibration
method in the Supervisor level calibration setup menu.
On pressing the [Cal] key the display will show the busy symbol and a line of 7 dashes
and then after a few seconds will display 'CALIBRA
dashes will reappear, followed by 'CAL On
the busy symbol and a line of dashes. Finally 'CAL OFF
beep and the busy symbol and a line of dashes. A final beep will sound the end of
calibration and the display should return to '0 0 0 0 .000
complete and normal operations may proceed.

6.2.3 Calibration using External Calibration mass

Note: The calibration mass used should be a known accurate item, ideally with an
OIML or ASTM/ANSI classification appropriate to the accuracy of the balance.
On pressing the [Cal] key, the display will show the balance setting a new Zero condition
LOAd
LOAd
LOAd
LOAd
0 0 0 0
by showing "
The display will show the busy symbol and a line of dashes, and then after a few seconds
will display the required calibration mass. For example, for a 213e model the display will
L L L L 100
100
100
100
be "
" where 100 g is the required calibration mass.
g
Place the selected mass on the balance and press the [Select] key to continue. The
balance will automatically continue. The display will show the busy symbol and a line of
dashes, and after calibration is complete it will sound a beep and display "UNLOAD
Remove the weight. Another beep will be heard confirming the unloading action. The
balance will display the busy symbol and a line of dashes for a few seconds and then
sound a beep and return to normal weighing.

6.2.4 Automatic Calibration

The balance will indicate the need for calibration when the balance has automatic
calibration enabled and the set pre-conditions for automatic calibration have been met.
© Adam Equipment Company 2015
CAL On'. Then 'CALIbrA
CAL On
CAL On
". Make sure the pan is empty, then press the [Select] key to continue.
g
CALIBRA'. Then the busy symbol and a line of 7
CALIBRA
CALIBRA
CALIbrA' will appear again, followed by
CALIbrA
CALIbrA
CAL OFF' will be displayed, followed by a
CAL OFF
CAL OFF
000 g' or similar. Internal calibration is now
000
000
12
3016612482 Rev1-Aug15
UNLOAD".
UNLOAD
UNLOAD