Troubleshooting - Kodak 2100 User Manual

Intraoral x-ray system
Hide thumbs Also See for 2100:
Table of Contents

Advertisement

Troubleshooting

Problem
Nothing lights up
Control unit does not
light up
No x-ray emission
X-ray emission works,
but exposure is too light
or completely white
X-ray emission works,
but exposure is too dark
5-4
Table 11. Troubleshooting
Cause
Unit is disconnected.
Fuse F1 is burned out or defective.
Circuit breaker is off.
Control unit is disconnected.
Fuse F1 is burned out or defective.
Control unit is defective.
Generator is cooling.
Radiology control key is defective.
Generator is positioned incorrectly.
Exposure time is too short.
Development time is too short.
Developer is too cold.
Developer is too old or diluted.
RVG mode is incorrectly selected.
Receptor is facing the wrong way.
Unit was incorrectly installed.
Exposure time is too long.
Development time is too long.
Developer is too hot.
Developer is too concentrated.
RVG/film mode is incorrectly selected. Verify your exposure settings. (Refer to
SM700_K2100_03_en
Solution
Connect the unit.
Replace the fuse.
Turn on the circuit breaker.
Connect the control unit.
Replace the fuse.
Call a qualified service technician.
Wait for the I01 message to disappear.
Call a qualified service technician.
Adjust the position of the generator.
Increase the exposure time.
Increase the development time. (Refer to
the development instructions.)
Heat the developer.
Replace with fresh developer.
Verify your exposure settings. (Refer to
the exposure procedure.)
Reposition the receptor.
Call a qualified service technician.
Decrease the exposure time.
Decrease the development time. (Refer
to the development instructions.)
Cool the developer.
Adjust the concentration or replace the
developer.
the exposure procedure.)
9/2007

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents