Specifications
Environmental Effects
Acoustical noise:
•
Less than 55 dB at 1.5 m during idle or standby
•
Less than 75 dB momentary at 1 m during normal operation
Power Requirements
A power cord is provided with this equipment. All countries must use an Agency-
approved power cord with a plug type suitable for the country of use. Contact a qualified
dealer for help.
Connect the equipment to a power source that is suitable for the voltage and current
ratings shown on the rating label. The single-phase power source, with grounding, must
be provided within 2.5 m (8.0 ft) of the imager.
The wire must be insulation-rated for 600 V (ac). A dedicated line is recommended.
Network Requirements
The imager receives digital images from medical imaging devices (modalities) over a 10/
100Base-T or 1000Base-T Ethernet Network.