Cleaning A Series 4Tev; Cleaning Procedure - Decagon Devices AquaLab 4TE Operator's Manual

Water activity meter
Table of Contents

Advertisement

AquaLab
6. Cleaning and Maintenance
sample chamber and sensors. Th e sample chamber consists of all
surfaces inside the red o-ring when the lid is closed.

Cleaning a Series 4TEV:

If cleaning an AquaLab Series 4TEV, follow the cleaning procedures
listed below being careful not to get cleaning solution or alcohol on
the capacitance sensor fi lter (see illustration on previous page). Re-
peated exposure of cleaning materials or contaminants to the fi lter
may cause inaccurate readings. If the fi lter appears to be contami-
nated, it may need to be replaced. (To replace the capacitance sensor
fi lter, use a tweezer or small knife blade to pry up the edge of the
fi lter, being careful not to disturb the sensor beneath. Discard the
soiled fi lter, then with clean hands, press a new fi lter into place.)

Cleaning Procedure:

Cleaning your AquaLab is a multi-step procedure which in-
volves washing, rinsing, and drying for each specifi c area as out-
lined below (refer to illustration at the beginning of this chap-
ter to identify the location of the components to be cleaned):
1. Cleaning the Sample Chamber
Note: Be extremely careful not to damage the fan blades (see illustra-
tion) when cleaning the chamber.
a. Remove any debris that may have collected within or around
the sample chanber.
b. Wrap a NEW Kimwipe around the end of the thin plastic
rod (spatula) and moisten it with isopropyl alcohol or Deca-
gon Cleaning Solution. Note: Do NOT dip a used Kimwipe
into your container of IPA or cleaning solution (the IPA or
cleaning solution will become contaminated).
c. WASH—Clean upper chamber, o-ring, and all surfaces of the
block within the o-ring. You may need to replace the Kim-
34

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Aqualab duoAqualab 4tev

Table of Contents