SOLUTION:
In order to begin your trial period for your AquaLab Series 4 instru-
ment, you will need to contact Decagon Devices for instructions on
how to activate your demo.
13. DUO PROBLEM:
Test was run with wrong model.
SOLUTION:
1) On the measurement screen, toggle to the correct model us-
ing the up and down arrow keys. Th e moisture content value
will be updated to correspond with the model selected.
2) If the correct model is not available, the model may not be
loaded on the instrument.
a. To determine which models are loaded on the instrument,
cycle to the menu tab, select Moisture Content and then the
loaded models will appear.
3) If the correct model is not available, load the appropriate
model using Aqualink RG Software. Th e AquaLab DUO can
hold a total of 100 models at any one time. You may need to
remove a model using the RG Software or use the delete op-
tion in the % moisture calibration menu before you can add
a new one. Any model that is removed from the instrument
with AquaLink RG will be stored and may be reloaded again
later.