8. Once the reading is complete, a screen will display the water
activity measured as well as the moisture content based on the
target model. Adjust the moisture content reading using the
up and down arrows until it matches the moisture content
value obtained from your reference method and click Save.
Note: If you chose to edit an existing model, pressing save updates the
model but keeps the same name. If you chose to create a new model,
pressing save will bring up an annotation screen where you will enter
the new name for the model. Pressing the cancel button will return you
to the Confi guration menu and cancel the moisture content adjustment.
9. Re-measure the sample again in normal sampling mode. It
should now read the corrected moisture content value you
provided in the previous step.
If your moisture content readings are still inconsistent, contact
Decagon by email at support@decagon.com or by phone at 509-
332-2756 (800-755-2751 in US and Canada) for further instruc-
tions. If you purchased your AquaLab Instrument from one of our
international distributors, please contact them for local service and
support.