Vega VEGAPULS 63 Operating Instructions Manual page 56

Radar sensor for continuous level measurement of aggressive liquids or with hygienic requirements
Hide thumbs Also See for VEGAPULS 63:
Table of Contents

Advertisement

10 Supplement
Ambient conditions
Ambient, storage and transport tempera-
ture
Process conditions
For the pressure stages of the vessel pressure, you also have to note the specifications on the type
label. Always the lowest value is applicable.
Pressure stage, depending on the process fitting
Ʋ Slotted nut according to DIN 11851
and 11864-2, hygienic fitting LA and
LB
vessel pressure depending on the process fitting
Ʋ flange up to 2" (from 300 lb)/DN 50
Ʋ Flange from 3" (from 300 lb), DN 80
Ʋ Tuchenhagen Varivent
Ʋ SMS
Vessel pressure relating to the flange
nominal pressure stage
Process temperature, depending on the material of the antenna encapsulation (measured on the
process fitting)
Ʋ TFM-PTFE encapsulated
Ʋ TFM-PTFE hygienically encapsulated
(low temperature version)
Ʋ PFA hygienically encapsulated
Ʋ PTFE hygienically encapsulated,
process seal FKM
Ʋ TFM-PTFE hygienically encapsulated,
process seal EPDM
Ʋ TFM-PCTFE hygienically encapsu-
lated
Vibration resistance
Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Cable entry/plug
12)
Ʋ Single chamber housing
Not suitable for SIL applications.
10)
Tested according to the guidelines of German Lloyd, GL directive 2.
11)
Depending on the version M12 x 1, according to DIN 43650, Harting, 7/8" FF.
12)
56
-40 ... +80 °C (-40 ... +176 °F)
PN 16
-100 ... 1600 kPa/-1 ... 16 bar (-14.5 ... 232 psig)
-20 ... 1600 kPa/-0.2 ... 16 bar (-2.9 ... 232 psig)
-100 ... 1000 kPa/-1 ... 10 bar (-14.5 ... 145 psig)
-100 ... 600 kPa/-1 ... 6 bar (-14.5 ... 87 psig)
see supplementary instructions manual "Flanges ac-
cording to DIN-EN-ASME-JIS"
-40 ... +150 °C (-40 ... +302 °F)
-170 ... +150 °C (-274 ... +302 °F)
10)
-40 ... +150 °C (-40 ... +302 °F)
-20 ... +130 °C (-4 ... +266 °F)
-40 ... +130 °C (-40 ... +266 °F)
-40 ... +150 °C (-40 ... +302 °F)
mechanical vibrations with 4 g and 5 ... 100 Hz
– 1 x cable gland M20 x 1.5 (cable: ø 5 ... 9 mm), 1 x
blind plug M20 x 1.5
or:
– 1 x closing cap M20 x 1.5; 1 x blind plug M20 x 1.5
or:
– 1 x closing cap ½ NPT, 1 x blind plug ½ NPT
or:
– 1 x plug (depending on the version), 1 x blind stopper
M20 x 1.5
11)
VEGAPULS 63 • Profibus PA

Advertisement

Table of Contents
loading

Table of Contents