Service - PIN
Info
VEGAPULS 63 • Profibus PA
6 Set up with the display and adjustment module PLICSCOM
•
Measured value presentation
•
Adjustment
•
Medium
•
Inner diameter of the standpipe (with standpipe versions)
•
Vessel form
•
Damping
•
Linearisation curve
•
Sensor-TAG
•
Displayed value
•
Display unit
•
Scaling
•
Current output
•
Unit of measurement
•
Language
The following safety-relevant data are not read out or written:
•
HART mode
•
PIN
•
SIL
Copy sensor data
Copy sensor data?
In this menu item, the PIN is activated/deactivated permanently.
Entering a 4-digit PIN protects the sensor data against unauthorized
access and unintentional modifications. If the PIN is activated perma-
nently, it can be deactivated temporarily (i.e. for approx. 60 min.) in
any menu item. The instrument is delivered with the PIN set to 0000.
PIN
Disable permanently?
Only the following functions are permitted with activated PIN:
•
Select menu items and show data
•
Read data from the sensor into the display and adjustment mod-
ule.
In this menu item the most important sensor information can be
displayed:
•
Instrument type
•
Serial number: 8-digit number, e.g. 12345678
Instrument type
Serial number
12345678
•
Date of manufacture: Date of the factory calibration, e.g. 24. March
2015