Symptom
Measurement data or
The instrument's backup batteries may
settings are not held
be low immediately after purchase or
in memory, and dis-
following a period of prolonged use.
appear immediately.
Turn the power of the instrument ON
to charge the backup batteries. Under
this condition, the batteries can be fully
charged in 25 hours.
Relative gloss value
In this case of the TASK mode, if the
is displayed as "G---"
MASK/GLOSS setting wasn't "
in the TASK mode.
I+E", relative gloss value is not calcu-
lated and is displayed as "G---".
Check Point
*
Troubleshooting
Action
The backup batteries have an expected
service life of approximately ten years.
If you find that the instrument fails to
retain data in memory even after the
batteries have been fully charged, how-
ever, then it is likely that your batteries
have reached the end of its life and re-
quires replacement. Note that you can-
not replace the batteries yourself. For
information, please contact a KONICA
MINOLTA
SENSING-authorized
service facility.
Change the MASK/GLOSS setting of
/
the TASK to "
/I+E", and redownload
*
the TASK from SpectraMagic (Ver.3.2
or higher; except for Ver.3.5) to the
CM-2600d/2500d.
E-111