Konica Minolta CM-2600d Instruction Manual page 111

Spectrophometer
Hide thumbs Also See for CM-2600d:
Table of Contents

Advertisement

MessageCorrective Action
ER13
ERROR IN A/D
ER17
INCORRECT CLOCK OPERA-
TION
ER24
WHITE CALIBRATION IN-
COMPLETED
ER25
WRONG APERTURE
ER27
ERROR IN CHARGING
ER30
NO DATA IN MEMORY
Symptom/Possible Cause
Failed during A/D conversion.
• Faulty A/D converter
• Break down of circuits relating
to the A/D converter
Clock IC is not working correct-
ly.
• Clock IF's power voltage drop
or breakdown of circuits relat-
ing to the clock IC.
White calibration has not been
performed.
The measurement area set in
<FILE> screen does not match
the lens position selector switch
on the instrument.
Charging for flashing of Xenon
lamp can't be completed
• Breakdown of charging circuit
• Battery voltage drop
Data has been lost since the
memory's backup battery is ex-
hausted.
Error Messages
Corrective Action
Turn the power OFF, and then
turn it ON again. If this message
keeps
appearing,
contact
KONICA MINOLTA SENSING
-authorized service facility.
Contact a KONICA MINOLTA
SENSING-authorized service fa-
cility.
White calibration must be per-
formed each time the instrument
is turned ON. White calibration
must also be performed when the
measurement area or specular
component setting is changed.
Change the measurement area
setting or the position of the
measurement
area
switch according to the area to be
measured.
Contact a KONICA MINOLTA
SENSING-authorized service fa-
cility.
Turn the power ON to charge the
memory's backup battery. The
backup batteries can be fully
charged in 25 hours when the
power of this instrument is
turned ON. After the batteries are
fully charged, data backup can be
stored
for
approximate
months.
a
selector
4.5
E-109