4.2.1 F1 - CALIBRATION

F1
To enter the calibration parameter, press the [Tare] key when "
" is displayed.
UnloAd
The display will instruct you to remove any weight from the scale by displaying "
"
Press the [Tare] key.
LoAd 1
05
The display will tell you to add the calibration weight to the scale: "
" "
"
kiLo5
"
"
Add the weight shown or press the
key to clear the value and enter another value
of weight. Wait for the stability indicator and then press the [Tare] key.
SPAn
PASS
SPAn
FAIL
The display will show "
" "
" if the calibration is OK or it will show "
" "
"
if calibration could not be completed correctly.
JP On
The display will show "
". Remove the calibration weight.
Remove the jumper and the scale will run a self-test and return to normal operation.

4.2.2 F2 - RESOLUTION

This function is not active on the CPB-M series.

4.2.3 F3 - CAPACITY

F3
To enter this parameter, press the [Tare] key when "
" is shown.
The display will show the current capacity in 3, 6, 15 or 30 kilograms.
Press the
key to change the value.
Press [Tare] to accept the displayed value.
NOTE: The scale must be originally built for the capacity selected. The 3kg unit
uses the 5kg load cells, 6kg unit uses 10kg load cells, the 15kg unit uses 20kg load
cells and the 30kg unit uses 35kg load cells.
CPB-SH-e-0721
12