F10 - Lvd Mode; Weighing Units [Kg / G] - KERN CPB 3K1M Service Manual

Counting balance
Table of Contents

Advertisement

Pin 2
The display will show "
"
Enter the new Pin number a second time.
Press the [Tare] key to accept.
dOnE
FAiL
The display will show "
" if the number was accepted or "
" if the numbers were
not the same.
The scale returns to the parameter menu.

4.2.10 F10 - LVD MODE

It stands for Low Voltage Detect mode. The Low voltage detection can be enabled or
disabled.
F10
To enter this parameter, press the [Tare] key when "
" is shown.
on
oFF
The display will show if the LVD is "
" or "
".
Press the
key to change the value.
on
Press [Tare] to accept "
".
LVD
on
NOTE: Only "
" "
" is allowed for the CPB-M scales.
Turn the power off, remove the jumper and restart the scale. The new settings are now
in operation.

4.3 WEIGHING UNITS [kg / g]

The current software of CPB was standardised to weigh in [kg].
There is a function to enable the scale to weigh in [g].
The scale has to be opened and a connecting jumper has to be fitted to "JTAG" pins 1
and 2 on the main board to enable the [g] display.
NOTE: The selection between both units is not possible for the CPB-M scales.