4.0 SERVICE PARAMETERS

Place your balance upside down
Remove seal
For verified balances, the jumper is on a pin
For adjustment the jumper must be set on both pins
Position jumper for adjustment
4.1
USING "0000" TO ENTER THE
After the jumper has been placed in the position shown, apply power to the scale.
The display will ask for a code number, "
Enter the number "
the calibration press [Tare] and to skip the calibration, press
displayed.
NOTE: If the PIN number has been changed by other users, the correct Pin number will
be required in place of 0000.
CPB-SH-e-0721
0000
" then press [Tare]. The display will show "
Bottom of the balance
CALIBRATION
Pn
" on the Weight Display.
Position
of seal
PARAMETER
F1
CAL
"
"
". To enter
F2
Pin
. "
" "
" will be
8