4.2
USING "2006" TO ENTER THE
After the jumper has been placed in the position shown, apply power to the scale.
The display will ask for a code number, "
2006
Enter the number
The displays will show the first parameter
parameters press the
Press the [Tare] key to enter a parameter.
To exit a parameter, press the
The Weight Display window will show the parameter number and the Piece Weight
window will show the word describing the function.
When a parameter is entered by pressing the [Tare] key, the displays will guide you
through the parameter selected and the options available.
The parameters available are:
F1
CAL
F2
rES
F3
CAPA
F4
init
ZEro
F5
rE
ZEro
F6
SCSivE tArE
F7
Ad
Count
F8
ZEro
F9
Pin
F10
LVd
ModE
CPB-SH-e-0721
Pn
Pn
when "
" is displayed and then press [Tare].
key.
key.
To enter the Calibration
Resolution selection, not valid for CPB-M units
Select Capacity
Initial Zero Range
Re-Zero Range
Successive Tare Enable
Display the A/D counts
Set zero method.
Set new PIN number
Set low voltage detect mode
SERVICE
PARAMETERS
" on the Weight Display.
"F1"
"CAL"
.
To scroll through other
11