Techniques for Improving Measurement Performance

This section describes several programming techniques that can improve speed and efficiency. Most, but not all, of these
techniques relate to SCPI program design.
"Turn off Display Updates" on page 31
"Use Binary Data Format instead of ASCII" on page 31
"Minimize the Number of Bus Transactions" on page 31
"Use USB or LAN Connection instead of GPIB" on page 32
"Minimize DUT/instrument Setup changes" on page 32
"Avoid Automatic Attenuator Setting" on page 32
"Avoid using RFBurst trigger for Single Burst Signals" on page 32
"Optimize GSM Output RF Spectrum Switching Measurement (N9071A Measurement Application)" on page 33
"To make Power Measurements on Multiple Bursts or Slots use CALCulate:DATA<n>:COMPress?" on page 33

Turn off Display Updates

When the instrument is being operated remotely, there is no need to display data on the instrument screen. Display
updates slow down the measurement, so measurement speed may be increased by switching off updates.
Send :DISPlay:ENABle OFF to turn off the display. In this case, data remains visible, but will no longer be updated.

Use Binary Data Format instead of ASCII

The ASCII data format is the instrument default, since it is easier for humans to read and is required by SCPI for *RST.
However, data input/output is faster using the binary formats.
:FORMat:DATA REAL,64 selects the 64-bit binary data format for all numerical data queries. (The REAL,32 format,
which is smaller and somewhat faster, should only be used if you do not require full data resolution. Some frequency data
may require full 64 bit resolution.)
If you are using a PC rather than UNIX, you may need to change the byte order to little-endian, by sending
:FORMat:BORDer SWAP. For details, see the "Programming the Analyzer" chapter of any X-Series Help file or Users &
Programmers Reference PDF.
When using the binary format, data is sent in a block of bytes prefixed by an ASCII header. For details of the block format,
see
"Block Program Data" on page

Minimize the Number of Bus Transactions

When you are using the GPIB bus for control of your instrument, each transaction requires driver overhead and bus
handshaking, so minimizing these transactions reduces the time used.
You can reduce bus transactions by sending multiple SCPI commands per transaction. See
Commands on the Same Line" on page
When making the same measurement multiple times with small changes in the measurement setup, use the READ
command, which is faster than using INITiate and FETCh.
When changing the frequency and making a measurement repeatedly, you can reduce transactions by sending the
optional frequency parameter with the READ query (for example, READ:<meas>? {<freq>}). These optional
parameters are not available in certain modes, such as Spectrum Analyzer or Phase Noise.
The CONFigure/MEASure/READ commands for some measurements allow you to send center frequency setup
information along with the command (for example, MEAS:PVT? 935.2MHz). This sets the Power vs. Time
measurement to its defaults, then changes the center frequency to 935.2 MHz, initiates a measurement, waits until it
is complete and returns the measurement data.
Techniques for Improving Measurement Performance
21.
21.
31
SCPI Programming Fundamentals
"Writing Multiple