The information contained in this document is subject to change without notice. Keysight Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
Page 3
N9038A MXE EMI Receiver. Please be sure to read this entire document before attempting to perform this upgrade. Prerequisites There is one prerequisite that must be met before installing Option RT1 in an N9038A MXE EMI Receiver. It is: Options:...
Page 4
Option RT1 - Real-Time Spectrum Analysis Upgrade What you will find in this Document Contents ........................... page 5 Tools Required........................page 5 Instrument Hardware Requirement Verification ............. page 5 Installation Instructions ....................page 6 License Key Retrieval and Installation................. page 6 Upgrade Front Panel Label Installation...............
Verify that Option B85 is installed in the instrument by pressing System, Show, System and verify that there is an entry for: N9038A-B85 Analysis Bandwidth, 85 MHz If N9038A-B85 is not found in the instrument do not proceed any further with this upgrade. Option N9038AK-B85 will need to be installed before proceeding. Installation Note N9038-90037...
6. Verify that the license keys have been installed by pressing System, Show, System and make sure that the following options are present: N9038A-RT1 Real-time analysis up to maximum BW, basic detection N9038A-FT1 Frequency Mask Trigger >15 us signal duration...
Page 7
Nameplate Label Location Instrument Software Verification / Installation The minimum instrument software revision to support Option RT1 in the N9038A MXE EMI Receiver is A.16.00. If the instrument does not have this revision or later the instrument software will need to be upgraded.
Option RT1 - Real-Time Spectrum Analysis Upgrade Operation Verification This procedure will verify the basic functionality of the Real-Time Spectrum Analysis application. 12.On the instrument press Mode, Real-Time Spectrum Analyzer. 13.Press FREQ Channel, Center Frequency, 50 MHz. 14.Press Input / Output, RF Calibrator, 50 MHz. 15.Press Peak Search.
Page 9
Adjustments and Performance Verification Manual Procedures Required None Automated Adjustments Required None Automated Performance Verification Testing Required None For assistance, contact your nearest Keysight Technologies Sales and Service Office. To find your local Keysight office access the following URL: http://www.keysight.com/find/assist Installation Note N9038-90037...