Error Messages - Jadever JWA series User Manual

Table of Contents

Advertisement

Error Message

【Error Messages】

Reasons
The CPU unable to read the
EEPROM
Unable to read the 3 points
calibration range
Zero Point is Too High
Zero Points is Too low
Unstable Zero Point
Battery for RTC is too weak
Overload (The weights on the pan
is more than +9e of the max.
capacity )
Unable to accumulate due to
contradiction in different unit . You
cannot accumulate this two
together .
The last accumulation is more than
the preset accumulation allowed . Press
Solutions
Return To The Manufacturer
Do the 3 points calibration
Make sure the pan is empty when
turn on the scale or perform the 3
points calibration.
Make sure the pan is on the scale or
perform the 3 points calibration.
(1)Make sure there is no winds or
vibration .
(2)Change the stability /vibration
parameter accordingly.
Replace a new battery or press
abort.
Take away the overloaded objects.
Press
key twice to clear the all the
accumulation or press
return to normal weighing mode .
key twice to clear the all the
accumulation or press
to normal weighing mode .
JWA WEIGHING SCALE
V02
to
key to
key to return
43