Calibration; Calibration Interval; Test Equipment Requirements - Pacific Power Source ASX series Operation Manual

Table of Contents

Advertisement

6

CALIBRATION

This section describes the calibration of the ASX-Series AC Power Source.
6.1

CALIBRATION INTERVAL

The ASX-Series Power Source requires calibration once every six months or after service has
been performed to the system.
6.2

TEST EQUIPMENT REQUIREMENTS

The test equipment listed below is required for calibration of the ASX-Series Power Source.
1.
2.
3.
4.
5.
Current requirement
10A/1φ
10A/1φ
26A/1φ
40A/1φ
10A/1φ
10A/1φ
30A/1φ
40A/1φ
80A/1φ
SECTION 6
CALIBRATION
Digital Voltmeter:
Frequency counter:
Digital Ammeter: 3½ Digit True-RMS responding 5000 Hz bandwidth, min.
(Alternate Approach: Current transformer used in conjunction with the DVM.)
Oscilloscope:
Load (Resistive):
5A/φ - 2φ
13A/φ - 2φ
13.3A/φ - 2φ
3.3A/φ - 3φ
3.3A/φ - 3φ
10A/φ - 3φ
13.3A/φ - 3φ
26A/φ 3φ
4½ Digit True-RMS responding 5000 Hz bandwidth, min.
5 digit counter, min.
(Optional)
Varies by model number
@ Voltage
Model Name
@ 132 VAC
Model 115-ASX
l-n
@ 150 VAC
Model 120-ASX
l-n
@ 135 VAC
Model 140-ASX
l-n
@ 132 VAC
Model 160-ASX
l-n
@ 132 VAC
Model 315-ASX
l-n
@ 150 VAC
Model 320-ASX
l-n
@ 135 VAC
Model 345-ASX
l-n
@ 132 VAC
Model 360-ASX
l-n
@ 135 VAC
Model 3120-ASX
l-n
82

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents