6.3
CALIBRATION PROCEDURE
The calibration procedure which follows verifies that system gains are set properly and that the
system performance, relative to output power capability is intact. Gains in various signal paths
within the controller are adjusted by the procedure of Paragraph 6.3.1. Output power capability of
the power source is tested by the procedure of Paragraph 6.3.2.
THIS EQUIPMENT CONTAINS HIGH ENERGY, LOW IMPEDANCE CIRCUITS!! LETHAL POTENTIALS ARE
CARE MUST BE EXERCISED WHEN SERVICING THIS EQUIPMENT IN ORDER TO PREVENT SERIOUS
VOLTAGE AT THE TERMINALS RESPONDS INSTANTLY WHEN THE OUTPUT IS ACTIVATED.
OBSERVE THE FOLLOWING WHEN SERVICE, MAINTENANCE, OR CALIBRATION ARE REQUIRED:
1)
REMOVE ALL JEWELRY FROM HANDS, ARMS AND NECK WHEN SERVICING THIS
EQUIPMENT. THIS PREVENTS THE POSSIBILITY OF SHORTING THROUGH THE JEWELRY AND
CAUSING BURNS OR ELECTROCUTION OF THE OPERATOR.
2)
WEAR SAFETY GLASSES WHEN SERVICING THIS EQUIPMENT TO PREVENT EYE INJURY
DUE TO FLYING PARTICLES CAUSED BY ACCIDENTAL SHORT CIRCUIT CONDITIONS.
3) DO NOT REMOVE ANY PANEL OR COVER WITHOUT FIRST REMOVING THE INPUT SERVICE
BY OPENING ALL CIRCUIT BREAKERS.
4)
SERVICE OTHER THAN EXTERNAL CLEANING SHOULD BE REFERRED TO
PERSONNEL AUTHORIZED BY THE FACTORY TO SERVICE THIS EQUIPMENT.

6.3.1 CALIBRATE CONTROLLER

The first step in system calibration is to calibrate the controller. The controller calibration
procedure is unique to the installed controller. Calibrate the controller as stated in the controller
manual.
UPC Controller - Refer to Section 8 of the UPC-Series Operation Manual.
CONTAINED WITHIN THE CABINET.
OPERATOR INJURY OR EQUIPMENT DAMAGE.
83
SECTION 6
WARNING
CALIBRATION