Maintenance; Maintenance Intervals; Cleaning; Replacing The Gel Face - Dräger Quaestor 6000 Instructions For Use Manual

Hide thumbs Also See for Quaestor 6000:
Table of Contents

Advertisement

Available languages

Available languages

Fault
Cause
No communica-
The USB cable is
tion with the PC
faulty
software is possi-
ble (the test unit
The test unit is
is not recog-
faulty
nized).
7

Maintenance

7.1 Maintenance intervals

Work to be carried out After use Annually
Test the gel face
Carry out an inspec-
1)
tion
Carry out a complete
2)
overhaul
1) Only to be carried out by trained experts.
2) Dräger performs complete overhauls on test units in order to
ensure the availability of test units and support the workflow in
respiratory protection workshops. Test units may be subject to
different levels of wear due to the differing use of the test
units, however Dräger strictly recommends that a complete
overhaul is carried out by no later than six years.

7.2 Cleaning

WARNING
Risk of electric shock!
Penetrating moisture can cause a short circuit.
► Before cleaning the device, pull the power plug.
If required, clean the test unit with a moist soft cloth and
dry it.
For information on suitable cleaning agents and
disinfectants and their specifications, see docu-
ment 9100081 at www.draeger.com/IFU.

7.3 Replacing the gel face

1. Remove the old gel face from the test head.
2. Moisten the inside of the new gel face with water.
3. To place the new gel face onto the test head, first
position the eye with the measuring point on the test
head, and then position the other eye. Press the gel
face against the test head.
4. Ensure that the edge of the gel face is flush around
the entire test head.
Instructions for use
|
Quaestor 6000/8000
Remedy
Use a new cable
Contact service
personnel
Every
6 years
X
X
X
5. Dräger recommends mounting a full face mask to
the test head until the moisture between the test
head and the gel face has evaporated (e.g.
overnight) and then commencing the test.
8

Transport

Only transport the test unit when it is switched off. The
test unit must not be used during transportation in a
vehicle.
For safety reasons, Dräger recommends lifting and
carrying the device with one hand on the handle and
the other hand on the high pressure port.
Do not dispose of the packaging! It must be used for
all servicing.
9

Storage

Store the test unit in a dry place. Avoid direct sunlight
and temperatures above 60 °C.

10 Disposal

This product must not be disposed of as house-
hold waste. This is indicated by the adjacent sym-
bol.
You can return this product to Dräger free of
charge. For information please contact the
national marketing organizations or Dräger.

11 Technical data

Dimensions with test
head
Maintenance
50 x 55 x 65 cm
|
19