Camera control unit (CCU) rear panel

Warning
Using this unit for medical purposes
This equipment's connectors are not isolated.
Do not connect any device other than one which
conforms to IEC60601-1.
When an information technology device or AV
device that uses an alternating current is
connected, current leakage may result in an
electric shock to the patient or operator.
If use of such a device is unavoidable, isolate its
power supply by connecting an isolation
transformer, or by connecting an isolator between
the connecting cables.
After implementing these measures, confirm that
the reduced risk now conforms to IEC60601-1.
Caution
Do not come into contact with the terminals of the
rear panel connectors and patients at the same
time.
Doing so may result in a generation of voltage
that can be harmful to patients if the unit is
malfunctioning. Always disconnect the power
cord before connecting and disconnecting
connectors.
12.
RED (R-gain) knob (page 28)
1.
Equipotential ground connector
Used to make an equipotential ground
connection.
2.
- (power) connector (page 23)
3.
3D-SYNC IN (3D-SYNC input)
connector,
3D-SYNC OUT (3D-
SYNC output) connector (BNC type)
(page 47)
4.
CAMERA connector (page 20)
REMOTE connector block
1
1.
Remote contact switch connector 1, 2
(stereo mini jack) (page 46)
2.
RS-232C connector (D-sub 9-pin) (page
48)
Part Names and Functions