Sony MCC-500MD Instructions For Use Manual page 4

Exmor
Hide thumbs Also See for MCC-500MD:
Table of Contents

Advertisement

unit causes interference (which can be
determined by unplugging the power
cord from the unit), try these measures:
Relocate the unit with respect to the
susceptible equipment. Plug this unit and
the susceptible equipment into different
branch circuit.
Consult your dealer. (According to standard
IEC60601-1-2 and CISPR11, Class B,
Group 1)
Important EMC notices for use in the medical environments
• The MCC-500MD needs special precautions regarding EMC and needs to be installed and put
into service according to the EMC information provided in this instructions for use.
• The portable and mobile RF communications equipment such as cellular phones can affect the
MCC-500MD.
Warning
The use of accessories and cables other than those specified, with the exception of replacement
parts sold by Sony Corporation, may result in increased emissions or decreased immunity of the
MCC-500MD.
Guidance and manufacturer's declaration-electromagnetic emissions
The MCC-500MD is intended for use in the electromagnetic environment specified below.
The customer or the user of the MCC-500MD should assure that it is used in such an
environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
4
Compliance
Group 1
The MCC-500MD uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
Class B
The MCC-500MD is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
Class A
purposes.
Complies
Electromagnetic environment-guidance