Guidance and manufacturer's declaration - electromagnetic immunity
The MCC-500MD is intended for use in the electromagnetic environment specified below. The
customer or the user of the MCC-500MD should assure that it is used in such as environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the MCC-500MD is used exceeds the
applicable RF compliance level above, the MCC-500MD should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such
as reorienting or relocating the MCC-500MD.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
6
IEC 60601
Compliance
test level
level
3 Vrms
3 Vrms
150 kHz to
80 MHz
3 V/m
3 Vrms
80 MHz to
2.5 GHz
Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the MCC-500MD, including cables,
than the recommended separation distance
calculated from the equation appliance to the
frequency of the transmitter.
Recommended separation distance
d = 1.2√P
d = 1.2√P 80 MHz to 800 MHz
d = 2.3√P 800 MHz to 2.5 GHz
Where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
a
survey,
should be less than the compliance
level in each frequency range.
Interference may occur in the vicinity of
equipment marked with following symbol:
b