Spectrophotometer read error.

Cause
The Spectrophotometer Sensor to
be loaded into the
Spectrophotometer Carriage is not
installed correctly.
The color measurement pattern
was not printed correctly.
The color measurement pattern is
out of alignment.
The performance of part of the
Spectrophotometer Unit may be
impaired.

Clean calibration white tile because it may be dirty.

Cause
The White Calibration Tile may be dirty.
Turn off the power, and reinstall by pushing the two connectors all the way into the
Spectrophotometer Sensor.
(See "Replacing the Spectrophotometer Sensor of the Spectrophotometer Unit
(option).")➔User's Guide
Check the color measurement pattern. If the printing is faint, clean the printhead and
perform the operation again.
(See "Cleaning the Printhead.")➔P.55
Pulling or lifting a printed color measurement pattern may cause the pattern to come
out of alignment. Perform the operation again taking care not to touch the color
measurement pattern.
Execute Feed Priority to improve the printing position of the pattern for color
measurement.
(See "Automatic Banding Adjustment.")➔P.100
If the same message is displayed again after performing the above operation, set
Pattern Size to Large on the Control Panel menu and perform the operation again.
(See "Menu Structure.")➔P.66
Alternately, it may be resolved by selecting Air Drying in Drying Method in the
Color Measurement Preferences dialog box in the Color Calibration Management
Console and selecting the Dry Separated Chart check box if it is off.
(See "Acquiring Color Measurement Data For Any Color Chart (Windows).")➔User's
Guide
(See "Acquiring Color Measurement Data For Any Color Chart (Mac OS).")➔User's
Guide
Turn off the printer and wait a while before restoring power, and then perform the
operation again.
If the same message is displayed again, contact your Canon dealer for assistance.
Remove, clean, and reinstall the White Calibration Tile.
(See "Cleaning the Spectrophotometer Unit (option) Parts.")➔P.56
If this message is still displayed after cleaning the White Calibration Tile, we
recommend adjusting the Spectrophotometer Sensor.
Contact your Canon dealer for more information on adjustment services.
(See "Replacing the Spectrophotometer Sensor of the Spectrophotometer Unit
(option).")➔User's Guide
Corrective Action
Corrective Action