Performing a Control Solution Test

The AgaMatrix WaveSense JAZZ System has 2 different levels
of control solution: Normal and High. Select the level of
control solution you want to use.
The control solution is used to:
[1] Ensure that your AgaMatrix WaveSense JAZZ Meter and
Test Strips are working together properly.
[2] Practice testing without having to use your own blood.
Control solution tests should be performed when you:
[1] First get your meter.
[2] Suspect that your meter or test strips are not working
properly.
[3] Think your test results are not accurate.
[4] Have dropped or damaged your meter or exposed your
meter to liquids.
[5] Are advised by your healthcare professional to do so.
Control solution should only be used for 90 days after first
opening the bottle or until the expiration date printed on the
label, whichever comes first.
18
8100-02301 Rev G Artwork, Guide, Owners, mmol/L, AgaMatrix, WaveSense Jazz SEC, UK
IMPORTANT: Count forward 90 days from the date you
open a new bottle of AgaMatrix Control Solution. This is
your discard date. Write the discard date on the AgaMatrix
Control Solution bottle.
Important AgaMatrix Control Solution information:
[1] Use only AgaMatrix Control Solution with your AgaMatrix
WaveSense JAZZ System.
[2] Always replace the cap on the control solution bottle
immediately after use.
[3] Do not add water or any other liquid to control solution.
[4] Control solution tests should be performed within the system
operating temperature range of 10 °C to 40 °C (50 °F to 104 °F).
How to perform a control solution test:
[1] Inserting an AgaMatrix WaveSense
JAZZ Test Strip: Insert a new AgaMatrix
WaveSense JAZZ Test Strip into the meter.
The meter will be activated when the test
strip is fully inserted.
[2] Meter Ready for Test: The meter
now displays the droplet symbol and the
word "Fill". The AgaMatrix WaveSense JAZZ
System is now ready for you to apply control
solution. You do not need to set a calibration
code for this meter.