[4] Applying Blood Sample
Bring Test Strip to Blood Sample: The meter now displays
the droplet symbol. This means your AgaMatrix WaveSense
JAZZ System is now ready for you to apply blood. Immedi-
ately bring the AgaMatrix WaveSense JAZZ Test Strip to the
blood sample.
The test strip fills from the
tip. Do not apply blood
on the top surface of the
test strip.
The AgaMatrix WaveSense JAZZ Test Strip acts like a sponge
and draws the blood into the test strip through the sample
area. The visual fill window of the AgaMatrix WaveSense JAZZ
Test Strip will turn red.
This is the
visual fi ll
window.
DO NOT PUT
BLOOD HERE.
Apply
blood here.
26
The entire
visual fi ll
window will
turn red.
8100-02301 Rev G Artwork, Guide, Owners, mmol/L, AgaMatrix, WaveSense Jazz SEC, UK
WARNING: If the visual fi ll window does not completely fi ll,
you may get an inaccurate result. Retest and ensure that
the visual fi ll window is completely full.
IMPORTANT:
[1] Remove the test strip from the blood sample as soon as
you hear the beep.
[2] Do not press the AgaMatrix WaveSense JAZZ Test Strip
against the test site.
[3] Do not scrape blood onto the AgaMatrix WaveSense
JAZZ Test Strip.
[4] Do not apply blood to the top side of the AgaMatrix
WaveSense JAZZ Test Strip.
[5] Do not apply blood to the AgaMatrix WaveSense JAZZ
Test Strip when the test strip is out of the meter.
[6] Do not put blood or foreign objects into the AgaMatrix
WaveSense JAZZ test strip port. Doing so may lead to
system failure.
[7] Do not apply more blood after beep.
[8] Carefully read the test results on the display before
making any treatment decisions.
27