Performing a Control Solution Test

The AgaMatrix WaveSense JAZZ System has 2 different levels
of control solution: Normal and High. Select the level of
control solution you want to use.
The control solution is used to:
[1] Ensure that your AgaMatrix WaveSense JAZZ Meter and
Test Strips are working together properly.
[2] Practice testing without having to use your own blood.
Control solution tests should be performed when you:
[1] First get your meter.
[2] Suspect that your meter or test strips are not working
properly.
[3] Think your test results are not accurate.
[4] Have dropped or damaged your meter or exposed your
meter to liquids.
[5] Are advised by your healthcare professional to do so.
Control solution should only be used for 90 days after first
opening the bottle or until the expiration date printed on the
label, whichever comes first.