[3] Applying Control Solution: Shake
the control solution bottle. Discard the
first drop of control solution and wipe
the bottle tip clean using a clean tissue
or paper towel. Dispense a second drop
onto a clean surface, such as an unused
resealable plastic bag.
Bring the tip of the test strip to the control solution sample.
The AgaMatrix WaveSense JAZZ System will automatically
detect if a test is performed with AgaMatrix Control Solution.
This is the visual fi ll window.
DO NOT PUT CONTROL
SOLUTION HERE.
Apply control
solution here.
When control solution is applied, the visual fill window on the
AgaMatrix WaveSense JAZZ Test Strip turns blue. Your meter
will beep once indicating that you should remove the tip of
the AgaMatrix WaveSense JAZZ Test Strip from the control
solution sample.
20
The entire
visual fi ll
window will
turn blue.
8100-02301 Rev G Artwork, Guide, Owners, mmol/L, AgaMatrix, WaveSense Jazz SEC, UK
On the meter display, the droplet symbol fills.
The scrolling bar and 1-2-3 indicates that the
meter is calculating your results.
[4] Viewing Control Solution Test Result:
The result will appear on the display and, if
the volume is enabled, your meter will beep
once. The date and the abbreviation "Ctrl"
will also appear on the display.
Compare the result of your control
solution test to the range printed on
the AgaMatrix WaveSense JAZZ Test
Strip vial label for the control solution
level (Normal or High) that you used.
Your control solution result should fall
within this range. The control solution
test result is stored in memory.
To deactivate the meter, remove the
used AgaMatrix WaveSense JAZZ
Test Strip. Dispose of used materials
properly.
If your control solution test results are
out of the range, repeat the test. If the
results from the control solution tests
continue to fall outside that range, do not use the AgaMatrix
WaveSense JAZZ System to test your blood glucose. Call
Customer Care.
Droplet
Symbol:
Normal: 4.4-6.7 mmol/L
High: 12.2-17.8 mmol/L
FOR EXAMPLE ONLY.
Refer to your test
strip vial label for
corresponding range.
21