Agilent Technologies 81101A Reference Manual page 83

50 mhz pulse generator
Hide thumbs Also See for 81101A:
Table of Contents

Advertisement

:SYST:SET
:SYSTem:SET
Set & Query
Block data
Not applicable
In query form, the command reads a block of data containing the
instrument's complete setup. The set-up information includes all
parameter and mode settings, but does not include the contents of the
instrument setting memories, the status group registers or the
:DISPlay[:WINDow][:STATe] The data is in a binary format, not
ASCII, and cannot be edited.
In set form, the block data must be a complete instrument setup read
using the query form of the command.
:SYST:VERS?
:SYSTem:VERSion?
Query
"1992.0"
Use this command to read the SCPI revision to which the instrument
complies.
:SYST:WARN?
:SYSTem:WARNing[:COUNt]?
Query
Not applicable
Use this command to read the number of warnings which are currently
active. Note that the warning status of voltage, current, time and
frequency are also summarized by bits in the QUESTionable Status
register.