:VOLT[1]
[:SOURce]:VOLTage[1][:LEVel][:IMMediate][:AMPLitude]
Set & Query
Numeric
V with engineering prefixes.
High = Offset +
Low = Offset –
With Offset, see
page 71
1.00 V
100 mVpp to 10.0 Vpp (values are valid from 50
Use this command to program the amplitude voltage of the output signal.
Note that to set the output levels in terms of voltage, you first have to
execute the [:SOURce]:HOLD VOLTage command to enable the
[:SOURce]:VOLTage subsystem.
The available voltage range is limited by the combination of:
• Specified current limits
• Actual output impedance setting :OUTPut:IMPedance
• Actual expected load impedance setting:
:OUTput:IMPedance:EXTernal
To set the amplitude voltage:
:HOLD VOLT
:VOLT 5V
70
Amplitude
2
Amplitude
2
Enable VOLTAGE subsystem
Set OUTPUT amplitude to 5 V
into 50 )