Agilent Technologies 81101A Reference Manual page 75

50 mhz pulse generator
Hide thumbs Also See for 81101A:
Table of Contents

Advertisement

:VOLT[1]:LIM:STAT
[:SOURce]:VOLTage[1]:LIMit:STATe
Set & Query
ON | OFF | 1 | 0
OFF
Use this command to switch the output limits on or off. When you switch
on the output limits, you cannot program the output levels beyond the
programmed limits, until you switch off the voltage limits. The limits
apply whether you program high/low levels or amplitude/offset levels.
You can switch the limits on and off in both the [:SOURce]:CURRent
and the [:SOURce]:VOLTage subsystems but the current and voltage
limits are not enabled/disabled independently. The voltage and current
limits are always enabled/disabled together.
To set and activate the high and low voltage limits:
:HOLD VOLT
:VOLT:LIM 3V
:VOLT:LIM:LOW 0V
:VOLT:LIM:STAT ON
:STATus:OPERation
This command tree accesses the OPERation status group. The
OPERation status group is not used by the instrument. Therefore, this
command tree is redundant.
• :STATus:OPERation[:EVENt]?
• :STATus:OPERation:CONDition?
• :STATus:OPERation:ENABle
• :STATus:OPERation:NTRansition
• :STATus:OPERation:PTRansition
Enable VOLTAGE subsystem
Set OUTPUT high level voltage limit to 3 V
Set OUTPUT low level voltage limit to 0 V
Switch on OUTPUT limits
75