Amplitude
Use this command to program the offset current of the OUTPUT signal.
Note that to set the OUTPUT levels in terms of current, you first have to
execute the [:SOURce]:HOLD CURRent command to enable the
:SOURce]:CURRent subsystem.
The available current range is limited by the combination of:
• Specified Voltage limits
• Actual OUTPUT Impedance setting :OUTPut:IMPedance
• Actual Expected Load impedance setting
To program the offset current of the output signal:
:HOLD CURR
:CURR:OFF 50MA
:CURR[1]:HIGH
[:SOURce]:CURRent[1][:LEVel][:IMMediate]:HIGH
Set & Query
Numeric
A with engineering prefixes.
Amplitude = High – Low
High – Low
Offset =
Low level
+10 mA (50
into 50 )
10 V Output (from high Z into short): –396 mA to 400 mA typical
3.8 V (from 50
into short –82 mA to 152 mA typical
Use this command to program the high level current of the OUTPUT
signal. Note that to set the OUTPUT levels in terms of current, you first
50
Enable CURRENT subsystem
Set OUTPUT offset to 50 mA
2