Agilent Technologies 81101A Reference Manual page 73

50 mhz pulse generator
Hide thumbs Also See for 81101A:
Table of Contents

Advertisement

:VOLT[1]:LOW
[:SOURce]:VOLTage[1][:LEVel][:IMMediate]:LOW
Set & Query
Numeric
V with engineering prefixes.
Amplitude = High – Low
High–Low
Offset =
2
With high level, see
–500 mV
–10.0 V to 9.9 V (50
Use this command to program the low level voltage of the OUTPUT
signal. Note that to set the OUTPUT levels in terms of voltage, you first
have to execute the [:SOURce]:HOLD VOLTage command to enable the
[:SOURce]:VOLTage subsystem.
The available voltage range is limited by the combination of:
• Specified current limits
• Actual OUTPUT impedance setting :OUTPut:IMPedance
• Actual expected load impedance setting
:OUTPut:IMPedance:EXTernal
To set the low level voltage:
:HOLD VOLT
:VOLT:LOW 500MV
page 72
into 50 )
Enable VOLTAGE subsystem
Set OUTPUT low level to 500mV
73

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents