:FREQ:AUTO
[:SOURce]:FREQuency[:CW|:FIXed]:AUTO
Event
ONCE
Not applicable
Use this command to measure the frequency at the CLK-IN connector. If
the CLK-IN connector is the selected pulse frequency source, you can
then read the measured value with :FREQ?
To measure and read the frequency at the CLK-IN connector:
:TRIG:SOUR EXT
:FREQ:AUTO ONCE
:FREQ?
:HOLD
[:SOURce]:HOLD
Set & Query
VOLTage | CURRent
VOLT
Use this command to enable either of the [:SOURce]:VOLTage or
[:SOURce]:CURRent subsystems.
You can control the signal levels of the instrument's OUTPUT in terms of
voltage or current.
Select ext CLK-IN as pulse trigger
Measure frequency at CLK-IN
Query pulse frequency