Level Specifications - Agilent Technologies 81101A Reference Manual

50 mhz pulse generator
Hide thumbs Also See for 81101A:
Table of Contents

Advertisement

Leading and trailing edges can be programmed independently within the
following ranges (Maximum ratio 1:20):
5ns

Level Specifications

Level Parameters
Source impedance:
Maximum external voltage:
Short circuit current:
Normal/complement:
ON/OFF:
Limits:
External Load compensation
For loads 50 , the actual load impedance can be entered to correct the
output values.
96
Agilent 81101A
selectable 50
± 1% typical or 1 k
±24 V
±400 mA
selectable
relay connect/disconnect output (HiZ).
high and low levels can be limited to protect the
DUT