Agilent Technologies 81101A Reference Manual page 24

50 mhz pulse generator
Hide thumbs Also See for 81101A:
Table of Contents

Advertisement

QUEStionable Status Group
Bit
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
The QUEStionable Status group is used to report warning conditions
amongst the voltage, current, pulse timing and frequency parameters.
Warnings occur when a parameter, although not outside its maximum
limits, could be causing an invalid signal at the output because of the
actual settings and uncertainties of related parameters.
24
QUEStionable
Voltage warning
Current warning
Time warning
Unused, always 0
Unused, always 0
Frequency warning
Unused, always 0
Unused, always 0
Unused, always 0
Unused, always 0
Unused, always 0
Unused, always 0
Unused, always
Unused, always 0
Unused, always 0
Always 0