Agilent Technologies 81101A Reference Manual page 37

50 mhz pulse generator
Hide thumbs Also See for 81101A:
Table of Contents

Advertisement

External Width
Set External Width mode using the :EWIDth[:STATe] command:
:ARM:EWIDth ON
This command disables the arm-trigger system. The arm-trigger system is
reenabled by switching OFF EWIDth mode.
Pulses
Set Pulses mode by setting the :TRIGger:COUNt to 1 so that a single
triggered pulse period is generated for every arm event. The trigger
source sets the pulse period:
:TRIGger:COUNt 1
:TRIGger:SOURce INTernal1 Pulse period from internal osc.
:DIGital:PATTern OFF
Pulse period Source
internal OSC.
internal PLL
CLK-IN
The internal PLL (INTernal2) cannot be used as :TRIGger:SOURce
(pulse period source) if it is already being used as ARM:SOURce
(triggering rate).
Note that in triggered pulses mode the pulse period source is not relevant
because a single pulse is generated for each ARM event.
Burst
Set Burst mode by setting the :TRIGger:COUNt to the burst length
required. The trigger source sets the pulse period for the pulses within
the burst.
:TRIGger:COUNt 16
:TRIGger:SOURce INTernal1 Pulse period from internal osc.
:DIGital:PATTern OFF
Switch on EXT WIDTH mode
Single pulse period per arm event
Disable pattern data.
:TRIGger SOURce
INTernal[1]
INTernal2
EXTernal2
Burst of 16 pulse periods
Disable pattern data
37

Hide quick links:

Advertisement

Table of Contents
loading